12-24W Medical Switching Power Adaptor Series Interchangeable medical adaptor Folding medical adapter
Telephone Bargaining
Input Voltage: 100~240V AC 50~60Hz
Input plug:Fixed / Interchangeable / Folding plug
Input current: 0.6A Max.
Output Voltage:3~48V(+-5% tolerance)
Output current:0.1A-5.0A
Ripple: 1% peak to peak maximum
Over voltage protection & Over current protection & Over Temperature protection & Short circuit protection.
Temperature coefficient: All out+-0.04%℃ maximum
Transient response:Maximum excursion of 4% or better on all models.Recovering to 1% of final value within 500uS after a 25% step load change.
Operating temperature:0℃~+40℃.
Storage temperature:-40℃~+85℃;.
Relative humidity:5% to 95% non-condensing.
Setup up / Hold up / Rise time:Typical: 3000mS / 16mS / 50mS at full load.
Line regulation:+-1% maximum for any input voltage change between input voltage range
Load regulation:+-5%variations from minimum to maximum output current.
Insulation resistance:50Mohm minimum from output to ground.
Withstand voltage:4300VDC from input to output
Mean time between failure:100,000 hours minimum at full load 25℃ ambient
Energy Saving (Level VI / Level 6)
EMC requirements:FCC CLASS B, EN55022 CLASS B.EN 55024,EN55011,EN55024 CLASS B.
Safety requirements:IEC60601-1-2 (A)UL:60601-1-2 (B)CSA:C22.2 NO.60601-1-2( C)EN60601-1-2
Warranty: 3- Years.
Industry Category | Electrical-Equipment-Supplies/Electrical-Plugs-Sockets |
---|---|
Product Category | |
Brand: | |
Spec: | |
Stock: | |
Origin: |